ABSTRACT:
A simple and inexpensive method was developed with high performance liquid chromatography with PDA detection for determination of Tolterodine. The chromatographic separations were achieved on (250×4.6 mm), 5.0 µm make: X-terra C18 column employing Acetonitrile: 0.1% v/v phosphoric acid in Water (35:65, v/v) as mobile phase with gradient programmed at flow rate 1.0 mL/min was chosen. The detector wavelength of 210 nm was employed.
Cite this article:
Tentu Nageswara Rao, Imad Hussain, Y. Prashanthi, T. B. Patrudu. Forced Degradation Study for Tolterodine by HPLC with PDA Detection. Asian J. Pharm. Ana. 2019; 9(2):77-81. doi: 10.5958/2231-5675.2019.00015.2