Development and Validation of HPLC Method for the Estimation of Metaxalone in Spiked Human Plasma

 

Prachi Gangurde, Sandeep Sonawane*, Sanjay Kshirsagar, Santosh Chhajed

Department of Pharmaceutical Analysis, MET’s Institute of Pharmacy, Bhujbal Knowledge City,

Adgaon, Nashik 422003

*Corresponding Author E-mail: sandeeps.iop@gmail.com

 

ABSTRACT:

A simple, rapid, accurate, precise and specific HPLC method was developed for the estimation of metaxalone (MAT) in spiked human plasma using darunavir (DAR) as an internal standard (IS). The drug and IS were successfully separated on BDS-C18 column using acetonitrile: water (50:50 %, v/v) as mobile phase in an isocratic mode at a flow rate of 1 mL/min. The MAT and DAR were extracted from the plasma by liquid-liquid extraction using ter-butyl methyl ether (TBME). Eluents were detected at 217 nm. The calibration curve was linear in the range of 0.1-4.0mg/mL. Weighted linear regression was applied with a weighting factor of 1/x to avoid heteroscedasticity. The method was validated for accuracy, precision, recovery, stability and for carry over as per US-FDA guidance for bioanalytical method validation, May 2018.

 

KEYWORDS: Metaxalone, human plasma, HPLC, bioanalytical method validation.

 

 


INTRODUCTION:

Chemically MAT (Fig 1) is, 5-[(3,5-dimethylphenoxy)methyl]-1,3-oxazolidin-2-one [1]. It is a skeleton muscle relaxantused in the short-term treatment of painful muscle spasms [2]. It isgenerally used alone or in combination with diclofenac. Several analytical methods have been reported for the estimation of MAT, includes estimation of MAT in pharmaceutical formulation by UV spectroscopy alone [3] and in combination with diclofenac [4,5,6], quantitation of MAT in different physiological fluids by UV spectroscopy [7], estimation of MAT in pharmaceutical formulations by HPLC [8,9,10,11] and estimation of MAT in combination with diclofenac by HPTLC [12].

 

Along with this, several analytical methods have been reported for the estimation of MAT in human plasma by LC-MS [13,14,15,16,17] and by high-throughput LCMS/MS [18] as well estimation of MAT in rat plasma (19) by LCMS and by LCPDA [20].

 

 

Fig1: Chemical structure of MAT

 

In the present paper, a simple and rapid HPLC-UV method has been describedfor the estimation of MAT in spiked human plasma. MAT was extracted from human plasma using liquid-liquid extraction with suitable organic solvent. Further, the method was validated as per US-FDA Bioanalytical Method Validation: Guidance for Industry May 2018 [21].

 

MATERIALS AND METHODS:

Chemicals and Reagents:

Pharmaceutical grade MAT was supplied as a gift sample from Amnel pharmaceuticals Pvt Ltd., Ahmedabad, India. Blank human plasma sampleswere procured as a gift sample from Dr. Vasantrao Pawar Medical College Hospital and Research Centre, Nashik, India. The pooled plasma sample was prepared by mixing thoroughly obtained plasma samples from six different sources. Methanol. acetonitrile used in the analysis were of HPLC grade and other chemicals were of analytical grade. All chemicals were purchased from Merck Ltd., Mumbai, India. Double distilled water used in the analysis was prepared freshly and used after filtration through Durapore 0.45m × 47 mm membrane filter papers, purchased from Milipore India Pvt Ltd., Bengaluru, India.

 

Instrumentation and chromatographic condition:

HPLC analyses were carried out using a JASCO HPLC system consists of dual PU-2080 plus pumps, UV-2075 UV multi-channel UV detector and equipped Rheodyne (7725i) injection system with 100mL sample loop. Data obtained from the system were collected and processed using Borwin Chromatography software (version 1.50).

 

All chromatographic separations were carried out on Phenomenex Hyperclone C18 (BDS) column (250 × 4.6 mm, 5m) using a mixture of acetonitrile: water (50:50%, v/v) as a mobile phase in an isocratic mode at a constant flow rate of 1mL/min. All eluents were detected at the 217nm, the absorption maxima of the drug.

 

Liquid-liquid extraction:

An aliquot of 1mL of pooled human plasma was taken in a stoppered glass tube of capacity 20mL. To this tube, 100mL of 100mg/mL of MAT, 100mL of 100mg/mL of IS and 200mL of 1% formic acid was added. The resulting solution was vortex mixed for 3min. Further, the extracting organic solvent, TBME was addedinto it and again vortex mixed for 5 min. The tubes were centrifuged at 4ºC, for 10min at 3000 rpm. The organic layer was transferred to another tube and the solvent was evaporated under the stream of nitrogen. The residue was further reconstituted with 500mL of the mobile phase and the 100mL was injected into the HPLC system. 

 

Preparation of calibration curve (CC) standards and Quality Control (QC) samples:

The standard stock solution of MAT was appropriately diluted with methanol to get six different working standard solutions having concentrations of 1, 2, 3, 4, 8, 10, 16, 20, 32 and 40mg/mL. To obtain the calibration curve (CC) standard, 1mL aliquots of the pooled blank plasma sample were taken in nine stoppered glass tubes of 20 mL capacity, in each tube, 100mL of working standard solution of MAT, 100mL of 100mg/mL of IS and 200mL of formic acid was added to get results CC standards of 0.1, 0.2, 0.3, 0.4, 0.8, 1.0, 1.6, 3.2 and 4.0 mg/mL. All CC standards were similarly processed as per the procedure described in liquid-liquid extraction and finally injected into the HPLC system under optimized chromatographic conditions in six replicates. The chromatograms of CC standards were processed to obtain the peak area of MAT and IS and the retention factor was calculated for each CC standard by calculating the ratio of area of MAT to the area of IS.

 

Along with the CC standards, QC samples consisting of 0.3mg/mL of LQC, 1.6mg/mL of MQC and 3.2mg/mL of HQC were prepared.

 

Selection of calibration model:

Data obtained from calibration experiments were subjected to unweighted and weighted linear regression with the weighting factor of 1/x, 1/Öx and 1/x2. From the tested regression models, the appropriate model was selected with the minimal % RE and heteroscedaticity in the linear range and was used for further calculations.

 

Validation:

The developed method was validated as per the US-FDA guidelines Bioanalytical Method Validation: Guidance for Industry, May 2018 with respect to selectivity, accuracy, precision, recovery, stability and carry-over.

 

Selectivity was carried out at lower limit of quantitation (LLOQ) at the concentration of 0.1mg/mL against blank response of plasma obtained from six different sources. Accuracy and precision were carried out by analysing and processing five replicates of LQC, MQC and HQC over five days. The accuracy was estimated as the mean % RE and precision was measured in terms of % RSD. Recovery was calculated by comparing peak areas of processed QC samples to the standard dilutions. Stability studies conducted includes, stability at room temperature, stability at -20°C, bench-top stability and freeze-thaw stability. The % nominal and % RSD were calculated. Further, the carry-over experiment was conducted by injecting a series of sample injections in order of blank solution, unextracted upper limit of quantitation (ULOQ), blank solution, extracted blank matrix, extracted ULOQ and extracted blank matrix.

RESULTS AND DISCUSSION:

Different compositions of mobile phases were tried to separate the MAT and IS from the plasma components. It was found that acetonitrile: water (50:50 %, v/v) gave adequate retention and resolution in an isocratic mode at a flow rate of 1mL/min at 217nm. The retention time for MAT and IS was found at 5.4min and 6.9min, respectively. Different organic solvents (ethyl acetate, chloroform, TBME and diethyl ether) were used to extract the MAT and IS from plasma samples. It was found that MAT and IS were well extracted with TBME compared to other organic solvents. Also, the recovery of MAT and IS was found to improved when plasma samples was acidified with 1 % formic acid along with TBME. Hence, it was decided to add 200mL of 1 % formic acid in each sample. The results of % recovery of MAT in different organic solvents are depicted in Table 1 and the chromatogram of blank plasma extracted with TBME and the representative chromatogram of MAT and IS extracted in TBME are shown in Fig 2 and Fig 3, respectively. Under mentioned liquid-liquid extraction, the recovery of MAT and IS were found to be 50.68 % and 43.99%, respectively. For the highest concentration of MAT (4mg/mL), when different concentrations of IS were analysed, it was found that 10mg/mL of IS gave the peak area between 30-60% to that of the highest CC standard peak area of MAT.

 

Table 1: Recovery of MAT and IS in different organic solvents

Organic solvent

% Recovery of MAT

% Recovery of IS

n-hexane

5.06

8.26

Chloroform

30.67

11.25

Dichloromethane

31.87

14.75

Ethyl acetate

34.84

15.26

Diethyl ether

37.64

17.56

Ter-butyl methyl ether (TBME)

40.63

20.56

TBME with 1 % formic acid

50.68

43.99


 


Fig. 2: Chromatogram of blank plasma extracted in TBME and 1% formic acid

 

Fig. 3: Representative chromatogram of MET and DAR extracted in TBME and 1% Formic acid extract (RT of MAT: 5.048 min and DAR (IS): 6.967 min)


 

The data from the calibration experiment revealed the increase in the standard deviation (±SD) with increasing concentrations (Table 2) suggests the need of weighted linear regression. When the obtained calibration data were subjected to weighted linear regression with weighting factor of 1/x, 1/Öx and 1/x2, it was observed that, the weighted linear regression with 1/x weighting factor shows minimal % RE and F values. Hence, it was decided to use a weighted linear regression with 1/x weighting factor for further calculations with the regression equation of y = 0.00014 + 0.01459x.

 

Table 2: Calibration Curve (CC) standard data of MAT

CC No.

Concentration in mg/mL

RF (mean ± SD) (n=6)

1

0.1

0.0278±0.0011

2

0.2

0.0404±0.0012

3

0.3

0.0615±0.0016

4

0.4

0.0757±0.0028

5

0.8

0.1131±0.0029

6

1.0

0.1578±0.0039

7

1.6

0.2425±0.0047

8

2.0

0.2895±0.0051

9

3.2

0.4760±0.0167

10

4.0

0.5645±0.01967

 

The method was selective at LLOQ concentration of 0.1 µg/mL, as the responses of the plasma samples from six different sources were found less than 20 % of the response obtained for LLOQ (Table 3).

 

Table 3: Blank responses and LLOQ peak areas of MAT

Sr. No.

Blank response (µV.sec)

Peak areas at LLOQ (µV.sec)

% peak area in blank

1

1140

17394

6.50

2

1053

13803

5.91

3

1004

26104

3.84

4

1215

27560

4.40

5

1025

21633

4.73

6

1016

19979

5.08

 

Also, the chromatogram of blank plasma extract does not show any significant peak at the retention times of MAT and IS. The results of accuracy and precision studies are shown in Table 4, the method was found accurate and precise as the obtained RE and %RSD were found within ± 15 and less than 15%.

 

 

Table 4: Results of accuracy and precision studies

QC Level

Conc. Added (mg/mL)

Intra-day

Inter-day

Mean conc. (n=1)

% RE

% RSD

Mean conc. (n=1)

% RE

% RSD

LQC

0.3

312.76

-3.05

6.5

284.99

-2.33

4.14

MQC

1.6

1596.41

-0.22

2.0

1570.68

-1.83

2.35

HQC

3.2

3141.35

-14.38

8.2

3002.72

-5.89

2.45

 

 

Table 5: Stability study o MAT at room temperature and at -20°C

QC Levels

Stability at Room Temperature

Stability at -20°C

% Nominal

% RSD

% Nominal

% RSD

2 h

4 h

6 h

2 h

4 h

6 h

10days

20days

30days

10days

20days

30days

LQC

99.96

95.83

100.50

1.17

6.1

2.8

99.41

95.74

99.00

1.40

8.18

3.23

HQC

100.50

101.30

100.10

1.22

1.70

4.06

99.06

99.72

100.30

2.64

1.39

1.17

 

 

 

Table 6: Bench-top and freeze thaw stability studies of MAT

QC Level

Bench-top stability

Freeze thaw stability

% Nominal

 

% RSD

% Nominal

% RSD

FT 1

FT 2

FT 3

FT 1

FT 2

FT 3

LQC

105.44

4.4

99.88

97.74

99.63

1.12

3.21

0.36

HQC

99.96

1.93

100.05

99.65

99.75

0.46

2.93

3.47

 

 


The results of stability studies concluded that the MAT remained stable in human plasma samples under the studied stability cycles. The results of stability studies are depicted in Table 5 and Table 6.

 

When carry-over study was conducted under mentioned sequence for MAT, it was observed that the response of mean extracted ULOQ was £ 20 % and £ 5 % of response of extracted IS. The results of carry-over test are presented in Table 7.

 

Table 7: Results for carry-over test

Sr. No.

Sample

Peak Areas (µV.sec)

1

Blank DI

0

0

2

AQ-ULOQ

669220.750

636385.00

3

BLANK DI

0

0

4

AQ-ULOQ

644226.500

615632.500

5

Blank DI

0

0

6

Blank plasma

0

0

7

Extracted ULOQ

300142.598

269848.355

8

Blank plasma

0

0

9

Extracted ULOQ

330272.250

342395

10

Blank plasma

0

0

 

ACKNOWLEDGEMENT:

Authors are thankful to the management and trustees of the Bhujbal Knowledge City, MET’s Institute of Pharmacy, Nashik for proving necessary analytical facilities, Dr. Vasantrao Pawar Medical College Hospital and Research Centre, Nashik for proving gift sample of blank human plasma and to Amnel Pharmaceuticals Pvt Ltd, Ahmedabad for proving gift sample of MAT. 

 

REFERENCES:

1.        http://www.chemspider.com/Chemical-Structure.14709.html (visited: Sept. 2018)  

2.        https://www.rxlist.com/skelaxin-drug.htm (visited: Sept. 2018)

3.        Babu KS, Kareemulla P, Reddy M, Ishq BM, Natrajan NB, Madhu M and Gopinath C.   Extractive Spectrophotometric Method for the Determination of Metaxalone in Bulk and its Pharmaceutical Formulation. Asian Journal of Research in Chemistry. 2013; 6(8): 905-907.

4.        Gabhane KB, Kasture AV, Shirkhande VN, Barde LN and Wankhede VP. Simultaneous spectrophotometric determination of metaxalone and diclofenac potassium in combined tablet dosage form. International Journal of Chemical Sciences.  2009; 7(1): 539-545.

5.        Patel CR, Kimbahune RV, Kabra PV, Harish AR and Nargund LVG. Spectrophotometric estimation of metaxalone and diclofenac potassium by multicomponent analytical method from tablet dosage form. Journal of Analytical and Bioanalytical Techniques. 2012; 3(3): 137-139.

6.        Ramya B, Vinod Kumar K, Ramalingam P and Reddy JR. UV derivative spectrophotometric method for simultaneous estimation of metaxalone and diclofenac potassium in combined dosage form. International Journal of Pharmaceutical Sciences and Research. 2012; 3(11): 4301-4305.

7.        Waman N, Ajage R, Kendre P and Kasture V. Development and validation of UV spectroscopy method for estimation of Metaxalone in bulk and microemulsion at different physiological PH. Indo American Journal of Pharmaceutical Research. 2014; 4(1): 101-106.

8.        Nagavalli D, Sankar ASK, Anandkumar K, Vetrichelvan T and Balaji M. Estimation of metaxalone in bulk and in tablet dosage form by RP-HPLC. Research Journal of Pharmacy and Technology. 2010; 3(2): 409-411.

9.        Praveena S, Muneer S, Famida B, Sowmya S, Ishaq BM and Ahad HA. Development of RP-HPLC Method for rapid Determination of metaxalone and in bulk and solid oral dosage form. International Journal of Chemical Sciences. 2014; 12(4): 1430-1438. 

10.      Sahu PK, Swain S, Prasad GS, Panda J and Murthy YL. RP-HPLC Method for determination of Metaxalone using Box-Behnken experimental design. Journal of Applied Biopharmaceutics and Pharmacokinetics. 2014; 2(2): 40-49.

11.      Sahu PK, Panda J and Swain S. Chemometric Assisted Ion-Pair Chromatography of Metaxolone and Diclofenac in Binary Mixture: A Mechanistic Study. Indian Journal of Pharmaceutical Education and Research. 2018; 52(2): 293-304.

12.      Rajput MP, Bharakar VV, Yadav SS, Mulla TS and Rao JR. Validated HPTLC method for simultaneous estimation of diclofenac potassium and metaxalone in bulk drug and formulation. International Journal of Comprehensive Pharmacy. 2011; 2(12): 1-4.

13.      Nirogi RV, Kandikere MS, Mudigonda K, Shrivastava W and Datla PV. Quantification of metaxalone in human plasma by liquid chromatography coupled to tandem mass spectrometry. Journal of analytical toxicology. 2006; 30(4): 245-251.

14.      Soni NR and Patel MB. Bioanalytical method development and validation of metaxalone in human plasma by LC-MS/MS. International Journal of Pharmaceutical Research and Allied Sciences. 2015; 4(3): 100-108.

15.      Sucharitha C, Goli KA, Saravanan G and Perumal DV. Quantization of Metaxalone in Human Plasma using High-Performance Liquid Chromatography-Tandem Mass Spectroscopy. Journal of Advanced Pharmacy Education and Research. 2015; 5(2): 70-76.

16.      Roy, SMN, Mangaonkar KV, Yetal SM and Joshi SS. LC-MS-MS method for determination of metolazone in human plasma. E-Journal of Chemistry. 2008; 5(3): 634-640.

17.      Goswami D, Saha A, Gurule S, Khuroo A, Monif T and Vats P. Metaxalone estimation in biological matrix using high-throughput LC–MS/MS bioanalytical method. Journal of Chromatography B. 2012; 902: 132-136.

18.      Zhao L, Li Q, Shu C, Wang K and Ding L. Quantitative determination of metaxalone in human plasma by LC-MS and its application in a pharmacokinetic study. Asian Journal of Pharmaceutical Sciences. 2016; 11(5): 668-672.

19.      Kandasamy K, Gowarda VS, Nammalvar H and Govindarajan AKS. Bioanalytical method development, validation and quantification of metaxalone in rat plasma by liquid chromatography tandem mass spectrometry.  Journal of Bioanalysis and Biomedicine. 2012; S6: 006.

20.      Rizwana I, Prakash KV and Mohan GK. Development and validation of RP-HPLC method with diode array detection for estimation of metaxalone in rat plasma. Pharmaceutical Methods. 2014; 5(2): 61-68.

21.      Bioanalytical Method Validation Guidance for Industry, U. S. Department of Health and Human Services, FDA, May 2018 (https://www.fda.gov/media/70858/download, visited: Dec. 2018).

 

 

 

 

 

Received on 01.08.2019                 Accepted on 20.09.2019

© Asian Pharma Press All Right Reserved

Asian J. Pharm. Ana. 2019; 9(4):210-214.

DOI: 10.5958/2231-5675.2019.00035.8