Bioanalytical
Method Validation
Amruta
S. Kadam*, Nayana V. Pimpodkar1, Puja S. Gaikwad2, Sushila D. Chavan.3
*Lecturer - College
of Pharmacy (D.Pharm) Degaon,
Satara, (MH) India- 415 004.
1Principal – College
of Pharmacy (D.Pharm) Degaon, Satara, (MH) India- 415
004.
2Lecturer - College
of Pharmacy (D.Pharm) Degaon, Satara, (MH) India- 415
004.
3Lecturer - College
of Pharmacy (D.Pharm) Degaon, Satara, (MH) India- 415
004.
*Corresponding
Author E-mail: kadamamu84@gmail.com
ABSTRACT:
The Bioanalytical method validation includes all of the procedures that
demonstrate a particular method used for a quantitative measurement of analytes
in a given biological matrix is selective, sensitive, reliable and reproducible
for the deliberate use. Measurement of drug concentrations in biological
matrices is an important aspect of medicinal product development for those
products containing new active substances and generic products. Such data may
be required to support new applications as well as variations to licensed drug
products. The results of toxic kinetic, pharmacokinetic and bioequivalence
studies are used to make critical decisions supporting the safety and efficacy
of a medicinal drug substance or product. It is therefore paramount that the
applied Bioanalytical methods used are well characterized, fully validated and
documented to a satisfactory standard in order to yield reliable results.
KEYWORDS: BMV-bioanalytical method
validation.
INTRODUCTION:
Preface to bioanalytical method validation:1,2
Evidence that the substance
being quantified is
the deliberate analyte. The concentration range over which the analyte will be
determined must be defined in the bioanalytical method, based on the evaluation
of actual standard samples over the range, including their statistical
variation. This defines the standard curve.
It is important to use a
sufficient number of standards to define adequately the relationship between
concentration and reaction.. The number of standards
to be used will be a function of the active range and nature of the
concentration-response relationship. In many cases5 to8 concentrations
(excluding blank values) may define the standard curve. More than this
standards concentrations may be necessary for nonlinear than for linear
relationships.
Published guidelines:
Till
now there are various guidelines published, which gives us attention about
analytical method validation these are as follows,
·
ICH-Q2A “Text on Validation of Analytical
Procedure:(1994)
·
ICH-Q2B “Validation of Analytical Procedures:
Methodology: (1995)
·
CDER “Reviewer Guidance: Validation of Chromatographic
Method” (1994)
·
CDER “Submitting Samples and Analytical Data for
Method Validations” (1987)
·
CDER Draft “Analytical Procedures and Method
Validation” (2000)
·
CDER “Bioanalytical
Method Validation for Human
Studies” (1999)
·
USP<1225> “Validation of Compendia Methods” (current revision)
Typical analytical performance
characteristics used in method validation:
·
Specificity (Selectivity)
·
Linearity
·
Range
·
Accuracy
·
Precision
·
Detection Limit
·
Quantitation Limit
·
Robustness
·
System Suitability Testing
The
definitions by ICH guidelines are as follows, 3
Table no. 1
Validation characteristics |
ICH definition |
Accuracy |
The accuracy of an
analytical procedure expresses the closeness of agreement between the value
which is accepted either as a conventional true value or an accepted
reference value and the value found. This is sometimes termed trueness. |
Specificity |
Specificity is the ability
to assess unequivocally the analyte in the presence of components which may
be expected to be present. Typically these might
include impurities, degradants, matrix, etc. |
Precision |
The precision of an
analytical procedure expresses the closeness of agreement (degree of scatter)
between a series of measurements obtained from multiple sampling of the same
homogeneous sample under the prescribed conditions. Precision may be
considered at three levels: repeatability, intermediate precision and
reproducibility. |
Repeatability |
Repeatability expresses the
precision under the same operating conditions over a short interval of time.
Repeatability is also termed intra-assay precision . |
Intermediate precision |
Intermediate precision
expresses within-laboratories variations: different days, different analysts,
different equipment, etc. |
Reproducibility |
Reproducibility expresses
the precision between laboratories (collaborative studies, usually applied to
standardization of methodology). |
Limit of detection |
The detection limit of an
individual analytical procedure is the lowest amount of analyte in a sample
which can be detected but not necessarily quantitated as an exact value. |
Limit of quantitation |
The quantitation limit of
an individual analytical procedure is the lowest amount of analyte in a
sample which can be quantitatively determined with suitable precision and
accuracy. The quantitation limit is a parameter of quantitative assays for
low levels of compounds in sample matrices, and is used particularly for the
determination of impurities and/or degradation products. |
Linearity |
The linearity of an
analytical procedure is its ability (within a given range) to obtain test
results which are directly proportional to the concentration (amount) of
analyte in the sample. |
Range |
The range of an analytical
procedure is the interval between the upper and lower concentration (amounts)
of analyte in the sample (including these concentrations) for which it has
been demonstrated that the analytical procedure has a suitable level of
precision, accuracy and linearity. |
Robustness |
The robustness of an
analytical procedure is a measure of its capacity to remain unaffected by
small, but deliberate variations in method parameters and provides an
indication of its reliability during normal usage. |
The validation individuality
should be investigated based on the nature of the analytical method. outcome
for each applicable validation characteristic are compared against the selected
approval criteria and are summarized in the analytical method validation
report. ICH also provides recommendations on statistical analysis required to
demonstrate method appropriateness.
Reference
Standards:2,4,7.
Investigation of drugs and
their metabolites in a biological matrix is performed using calibration standards and quality control samples
(QCs) spiked with reference standards. The purity of the reference standard
used to prepare spiked samples can affect study data. For this reason,
legitimated analytical reference standards of known identity and purity should
be used to prepare solutions of known concentrations. If possible, the
reference standard should be identical to the analyte. When this is not
possible, an established chemical form (free base or acid, salt or ester) of
known purity can be used. Three types of
reference standards are usually used:
(1) Licensed reference
standards (e.g., USP compendial standards),
(2) Commercially-supplied
reference standards obtained from a trustworthy commercial source, and/or
(3) Other materials of
predictable purity custom-synthesized by an analytical laboratory or other
noncommercial establishment.
The source and lot number,
ending date, certificates of analyses when available, and/or internally or
externally generated evidence of identity and purity should be furnished for
each reference and internal standard (IS) used. If the reference or internal
standard expires, stock solutions made with this lot of standard should not be
used unless purity is re-established.
Selectivity:
Selectivity is the ability of
an analytical method to separate and measure the analyte in the crowd of other components in the
sample. facts should be provided that
the substance quantified is the intended analyte. Analyses of blank samples of the appropriate
biological matrix (plasma, urine, or other matrix) should be obtained from at
least six sources. Each blank sample should be tested for hindrance, and
selectivity should be ensured at the lower limit of quantification (LLOQ). Potential interfering substances in a
biological matrix include endogenous matrix components; metabolites; breakdown
products; and, in the actual study, concomitant medication and other
xenobiotics. If the method is planned to quantify more than one analyte, each
analyte should be tested to ensure that there is no obstruction.
Accuracy,
Precision, and Recovery:
The accuracy of an analytical
method describes the closeness of mean test results obtained by the method to
the actual value (concentration) of the analyte. Accuracy is determined by
repeat analysis of samples containing known amounts of the analyte (i.e., QCs).
Accuracy should be measured using a minimum of five determinations per
concentration. A minimum of
three concentrations in the range of expected study sample
concentrations is recommended. The mean value should be within 15% of the
nominal value except at LLOQ, where it should not deviate by more than 20%. The
deviation of the mean from the nominal value serves as the measure of accuracy.
The precision of an analytical
method describes the closeness of individual measures of an analyte when the
procedure is applied repeatedly to multiple aliquots of a single homogeneous
volume of biological matrix. Precision should be measured using a minimum of
five determinations per concentration. A minimum of three concentrations in the
range of expected study sample concentrations is recommended. The precision
determined at each concentration level should not exceed 15% of the
coefficient of variation (CV) except for the LLOQ, where it should not exceed
20% of the CV. The recovery of an analyte in an assay is the detector response
obtained from an amount of the analyte added to and extracted from
the biological matrix, compared to the detector response obtained for the true
concentration of the analyte in solvent. Recovery pertains to the extraction
efficiency of an analytical method within the limits of variability. Recovery
of the analyte need
not be 100%, but the extent of recovery of an analyte and of the
internal standard should be consistent, precise, and reproducible.
Calibration
Curve:
A calibration (standard) curve
is the relationship between response and known concentrations of the analyte.
The relationship between response and concentration should be continuous and
reproducible. A calibration curve should be generated for each analyte in the
sample. The calibration standards can contain more than one analyte. A
calibration curve should be prepared in the same biological matrix as the
samples in the intended study by spiking the matrix with known concentrations
of the analyte. In rare cases, matrices may be difficult to obtain (e.g.,
cerebrospinal fluid). In such cases,
calibration curves constructed in stand in matrices should be reliable.
Concentrations of standards should be chosen on the basis of the concentration
range expected in a particular study. A calibration curve should consist of a
blank sample (matrix sample processed without analyte or internal standard), a
zero sample (matrix sample processed without analyte but with internal
standard), and at least six non-zero samples (matrix samples processed with
analyte and internal standard) covering the expected range, including LLOQ.
Method validation experiments should include a minimum of six runs conducted
over several days,
with at least four concentrations (including LLOQ, low, medium, and high)
analyzed in replica in each run.
a.
Lower Limit of Quantification (LLOQ)
Lower Limit of Quantification
(LLOQ) Contains The lowest standard on the calibration curve, should be
accepted as the LLOQ if the following conditions are met:
·
The analyte response at the LLOQ should be at least
five times the response compared to blank response.
·
Analyte peak (response) should be expressable,
discrete, and reproducible, and the back-calculated concentration should have
precision that does not exceed 20% of the CV and accuracy within 20% of the
nominal concentration. The LLOQ should not be confused with the limit of
detection (LOD) and/or the low QC sample.
·
The LLOQ should be recognized using at least five
samples and determining the CV and suitable confidence interval should be
determined.
b. Upper Limit of Quantification (ULOQ):
The highest
standard will define the ULOQ of an analytical method.
·
Analyte peak (response) should be reproducible and the
back-calculated concentration should have precision that does not exceed 15% of
the CV and accuracy within 15% of the nominal concentration
c. Calibration Curve/Standard
Curve/Concentration-Response:
·
The simplest model that adequately describes the
concentration-response relationship should be used. Selection of weighting and
use of a complex regression equation should be justified. Standards/calibrators
should not deviate by more than 15% of nominal concentrations, except at LLOQ
where the standard/calibrator should not deviate by more than 20%.
·
The acceptance criterion for the standard curve is
that at least 75% of non-zero standards should meet the above criteria,
including the LLOQ. Excluding an individual standard should not change the
model used. Exclusion of calibrators for
reasons other than failing to meet acceptance criteria and assignable causes is
discouraged.
d. Quality Control Samples (QCs):
·
At least three concentrations of QCs in duplicate
should be incorporated into each run as follows: one within three times the
LLOQ (low QC), one in the midrange (middle QC), and one approaching the
high end (high QC) of the range of the
expected study concentrations.
·
The QCs provide the basis of accepting or rejecting
the run. At least 67% (e.g., at least four out of six) of the QCs concentration
results should be within 15% of their theoretical values. At least 50%
of QCs at each level should be within 15% of their nominal concentrations. A confidence interval approach yielding
comparable accuracy and precision in the run is an appropriate alternative.
·
The minimum number of QCs should be at least 5% of the
number of unknown
samples or six total QCs, whichever is greater.
·
It is recommended that calibration standards and QCs
be prepared from separate stock solutions. However, standards and QCs can be prepared
from the same spiking stock solution, provided the stability and accuracy of
the stock solution have been verified. A single source of blank matrix may also
be used, provided absence of matrix effects on extraction recovery and
detection has been verified. At least one demonstration of precision and
accuracy of calibrators and QCs prepared from separate stock solutions is
expected. Acceptance/rejection criteria for spiked, matrix-based calibration
standards and QCs should be based on the nominal (theoretical) concentration of
analytes.
4.
Sensitivity:
Sensitivity is defined as the
lowest analyte concentration that can be measured with acceptable accuracy and
precision (i.e., LLOQ).
5.
Reproducibility:
Reproducibility of the method
is assessed by replicate measurements using the assay, including quality
controls and possibly incurred samples. Reinjection reproducibility should be
evaluated to determine if an analytical run could be reanalyzed in the case of
instrument interruptions.
6.
Stability:
The chemical stability of an
analyte in a given matrix under specific conditions for given time intervals is
assessed in several ways. Pre-study stability evaluations should cover the
expected sample handling and storeroom conditions during the conduct of the
study, including conditions at the experimental site, during delivery, and at
all other secondary sites. Drug constancy in a biological fluid is a function
of the storage situation, the physicochemical properties of the drug, the
matrix, and the container system. The stability of an analyte in a particular matrix
and container system is relevant only to that matrix and container system and
should not be extrapolated to other matrices and container systems. constancy
testing should evaluate the stability of the analytes during sample collection and handling, after
long-term (frozen at the intended storage temperature) and short-term (bench
top, room temperature) storage, and after freeze and melt cycles and the
analytical process. Conditions used in stability experiments should reflect
situations likely to be encountered during actual sample handling and
analysis. If, during sample analysis for
a study, storage conditions changed and exceeded the sample storage conditions
evaluated during method validation, stability should be recognized under these
new conditions.
The procedure should also
include an evaluation of analyte stability in stock solution. All stability
determinations should use a set of samples prepared from a freshly made stock
solution of the analyte in the appropriate analyte-free, interference-free
biological matrix. Stock solutions of the analyte for stability evaluation
should be prepared in an appropriate solvent at known concentrations. Stability
samples should be compared to freshly made calibrators. At least three
replicates at each of the low and high concentrations should be assessed.
Stability sample results should be within 15% of nominal concentrations.
a. Freeze and Thaw(melt) Stability:
During freeze/thaw stability
evaluations, the freezing and thawing of stability samples should mimic the intended sample
handling conditions to be used during sample analysis. Stability should be assessed for a minimum of
three freeze-thaw cycles.
b. Bench-Top Stability:
Bench top stability experiments
should be designed and conducted to cover the laboratory managing conditions
that are expected for study samples.
c. Long-Term Stability:
The storage time in a
long-term stability evaluation should equal or exceed the time between the date of first sample
collection and the date of last sample analysis.
d. Stock Solution Stability:
The stability of stock
solutions of drug and internal standard should be evaluated. When the stock solution exists in a different
state (solution vs. solid) or in a different buffer composition (generally the
case for macromolecules) from the licensed reference standard, the stability
data on this stock solution should be generated to rationalize the duration of stock solution storage stability.
e. Processed Sample Stability:
The stability of processed
samples, including the resident time in the auto sampler, should be determined
·
New technologies or
advancements:2,7
The Dried Blood Spot (DBS)
methodology has been successful in individual cases, the method has not yet
been widely accepted. Benefits of DBS
include reduced blood sample volumes collected for drug analysis and ease of
collection, storage, and transportation. A complete validation will be
essential prior to using DBS in synchronized studies. This validation should address,
at a minimum, the effects of the following issues: storage and handling
temperature, homogeneity of sample spotting, hematocrit, stability, and
carryover. Correlative studies with conventional sampling should be conducted
during drug development.
Statistics3,5,6,7
in Analytical Method Validation:
Statistical investigation of
data obtained during a method validation should be performed to reveal validity
of the analytical method. The statistics required for the interpretation of
analytical method validation results are the computation of the following-
1. mean,
2. Standard deviation,
3. Relative standard
deviation,
4.confidence intervals, and
5. Regression analysis.
These calculations are
classically performed using statistical software packages such as Excel,
Minitab, etc. The purpose of statistical analysis is to review a collection of
data that provides an understanding of the examined method characteristic. The
approval criterion for each validation characteristic are typically around the
individual values as well as the mean and relative standard deviation.
Mean:
Mean or average of an numbers
set is the essential and the most frequent statistics used. The mean is
calculated by adding all data points and dividing the sum by the number of samples.
It is typically denoted by x̄ (x bar) and is computed using
the following formula:
where Xi
are individual values and n is the number of individual data points.
Standard
Deviation:
The standard divergence or the
deviation of a data set is the evaluation of the spread of the values in the
sample set and is computed by measuring the variation or difference between the
mean and the individual values in a set. It is computed using the following
formula:
Where, Xi
is individual value, X̄ is the sample mean,
and n is the number of individual data points.
The relative standard
deviation is computed by taking the standard deviation of the sample set
multiplied by 100% and dividing it by the sample set average. The relative standard
deviation is expressed as percent. Typically, the acceptance criterion for
accuracy, precision, and repeatability of data is expressed in % RSD:
Confidence
Interval:
Confidence intervals are used
to designate the trustworthiness of an estimation. Confidence intervals gives
us limits around the sample mean to predict the range of the true population of
the mean. The prediction is usually based on possibility or probability of 95%.
The confidence interval depends on the sample standard deviation and the sample
mean. Confidence interval for:
Where, s is the sample deviation, X̄ is the sample mean, n is the
number of individual data points, and z is constant obtained from
statistical tables for z. The value of z
depends on the confidence level listed in statistical tables for z.
For 95%, z is 1.96. For small samples, z can be replaced by
t-value obtained from the Student’s t-distribution tables. The value of t
corresponds to n-1. Following Table provides an example of a classic data
analysis summary for the estimation of a system correctness or precision for a
high-powered liquid chromatography (HPLC) analysis.
Table no. 2:An Example of a System Precision Determination for a
HPLC Analysis.
Injection Number |
Area Response for analyte peak in standard |
1 |
451662 |
2 |
450752 |
3 |
447638 |
4 |
452541 |
5 |
449321 |
6 |
448747 |
Mean |
450110 |
Standard Deviation |
1861 |
%RSD |
0.41 |
95% confidence interval |
448621 to 451599 |
The calculated
confidence interval in Table indicates that the range of the true population of
the mean is between 448621 and 451599.
Regression analysis is used to evaluate a linear
relationship between test results. A linear relationship is usually evaluated
across the range of the analytical practice. The data obtained from analysis of
the solutions prepared at a range of different concentration levels is usually
investigated by plotting on a graph. Linear regression evaluates the
relationship between two variables by fitting a linear equation to experimental
data. A linear regression line has an equation of the form-
Y = bo
+ b1X,
Where,
X-
Independent or self-determining variable and Y - Dependent
variable.
The slope of the line is b1,
and bo is the intercept
(the value of y when x = 0). The statistical
procedure of ruling the “best-fitting” straight line is to obtain a line
through the points to reduce the deviations of the points from the approaching
line. The best-fit criterion of integrity of the robustness or fitness is known
as the principle of least squares. Following Table provides an example of data
that is evaluated for linearity.
Table no. 3:Solution
Concentration versus Measurements:
Concentration X |
Response Y |
Measurement |
|
10 |
213 |
20 |
378 |
30 |
629 |
40 |
848 |
50 |
994 |
60 |
1227 |
In this example, measurement
values (Response Y) are plotted against corresponding concentration (X), refer
to Figure
Figure no.-1
In this case, the data
obviously shows a linear relationship. The fitted or estimated regression line
equation is computed using the subsequent formula:
Y = b0 + b1X + ei
Summary
or Outline of work:
Measurement of drug
concentrations in biological matrices (such as serum, plasma, blood, urine, and
saliva) is an vital attribute of medicinal product
development.
From this review we came to
know regarding these validation parameters, and off course how important those
are. It is therefore vital that the applied Bioanalytical methods used are well
characterized, completely validated and documented to an adequate standard in
order to yield consistent or reliable results.
REFERENCES:
1. Rama Rao Kalakuntla and K. Santosh Kumar, Bioanalytical Method Validation: A Quality
Assurance Auditor View point, Journal of Pharmaceutical Sciences and Research, Vol.1(3), 2009,pg.
no.1-10.
2.
U.S.
Department of Health and Human Services Food and Drug Administration Center for
Drug Evaluation and Research (CDER) Center for Veterinary Medicine
(CVM)September 2013 Biopharmaceutics Revision 1.
3.
Eugenie Webster (Khlebnikova) ,Statistical Analysis in Analytical
Method Validation, IVT
Dec 16, 2013.
4.
Vinod
P. Shah,1 Kamal K. Midha, Bioanalytical Method
Validation—A Revisit with a Decade of Progress, Pharmaceutical Research, Vol. 17, No. 12, 2000.
5.
http:www.authorstream.com.
6.
http://en.wikipedia.org/wiki/bioanalysis.
7.
www.fda.gov/downloads/drugs/Guidelinescompliance
Regulatory information
/Guidance/ucm368107.pdf.
Received on 16.11.2015 Accepted
on 28.12.2015
© Asian Pharma Press All Right Reserved
Asian J. Pharm. Ana. 5(4): October- December,
2015; Page 219-225
DOI:
10.5958/2231-5675.2015.00035.6