Platelet Count Estimation by Peripheral
Smear Method and Automated Method in Pregnant Women
Dr. V. Kavitha,
B. Srinivas
Asst. Professor of Obstetrics and Gynecology, CKM Hospital,
Warangal-506002
*Corresponding Author E-mail: bethikavitha@yahoo.com
ABSTRACT:
Background
Platelet
count is an important investigation done in pregnant women. Platelet count is routinely done by
automated method. The automated cell counters are not available at all hospital
setups especially
in rural side. Platelets can also be estimated from the peripheral smears,
which can be easily done at any set up.
Aims and Objective: This study was conducted to compare the platelet estimation by
peripheral smear method and automated method.
Materials and Methods: Platelet estimation was done in 50 normal pregnant women by
stained peripheral smear and automated method. Platelet counts were expressed
in Mean ± SD. Statistical analysis was done by student’s t test.
Results:
Platelet counts were 2.76 ± 0.71 and 2.64 ± 0.73 lacs/mm3 by peripheral smear and automated method
respectively with p value 0.4.
Conclusion: There was no significant difference between two methods, hence it proves that the two methods are same.
KEYWORDS: Platelet Count; Peripheral Smear;
Automatic Analyser
INTRODUCTION:
Platelet count
is routinely advised in pregnant women. Thrombocytopenia has been more commonly
diagnosed in
pregnant women in the last 20yrs because platelet counts are included with the
automated blood cell counters.[1] There are several causes of
thrombocytopenia in pregnancy. The most common cause for thrombocytopenia in
pregnancy are gestational thrombocytopenia and
pregnancy induced hypertension (PIH). The degree of thrombocytopenia increases
with severity of disease.[2] Lower the
platelet count, greater are maternal and fetal morbidity and mortality.[3]
Platelet
count can be done by manual method or by automated analyzer. There are two
types of manual methods traditional method and alternate estimation. Traditional
method includes heamocytometry and stained peripheral
smear method. Alternate method is the average number of platelets per oil
immersion field (OIF) multiplied by the patient’s haemoglobin
concentration in g/dl and then multiplied by 1000 to yield platelet count
estimation per microliter.[4] Even though
automated cell counters are very sophisticated and accurate but they are not
available in all hospitals, particularly in rural side. This becomes a limiting
factor for doing platelet count as a routine investigation as a part of regular
antenatal checkups in rural areas. One of the manual methods which can be done
with minimal available equipment is the stained peripheral smear method. So we
wanted to study whether there is any difference in platelet count by these two
methods.
MATERIALS AND
METHODS:
50
second trimester normal pregnant women without history of hypertension or any
other systemic disorders which affect platelet count were recruited for the
study. The subjects were clinically examined. The capillary blood was drawn
under complete aseptic precautions, smears were prepared immediately and
stained using Leishman’s stain following standard
protocol. Platelets are counted in 10 oil immersion field.[5-7]
The average number of platelets is multiplied by 20,000 and the platelet
count is expressed as lacs/mm3. Data were
expressed in mean ± SD. Comparison between two methods was done by Student’s ‘t’ test. A
‘p’ values less than 0.05 were considered as significance.
RESULTS:
Platelet
count by peripheral smear method was 2.76 ± 0.71 lacs/mm3
and by automated method was 2.64 ± 0.73 lacs/mm3 with
p value of 0.4 (Table 1). There was no statistically significant difference
between two methods.
TABLE 1.PLATELET ESTIMATION BY TWO METHODS
|
Manual Method |
Automated Method |
P-Value |
Platelet estimation |
2.76 ± 0.71 |
2.64 ± 0.73 |
0.4 |
FIG1.Comparision
of platelet counts
DISCUSSION:
This
study was conducted to compare the platelet estimation by peripheral smear
method and automated method. There was no statistically significant difference
between two methods. Thus our results indicate that estimation of platelets by
peripheral smear method is simple, reliable, rapid, and cheaper which can be
performed even at the rural set up where there is no well-equipped laboratories.
This estimation can be helpful in assessing the severity of the disease and
early diagnosis of thrombocytopenia, so that the patients can be referred to
higher centers for the management as early as possible Gestational
thrombocytopenia is a benign common disorder, is the numeric platelet
deficiency seen most frequently in obstetrics. Second in frequency is the count
estimate per micro liter. Alternate estimation method was the average number of
platelets per oil immersion field multiplied thrombocytopenia characteristic of
HELLP syndrome a severe form of pre-eclampsia.[8] Hypertensive disorders account for 21%.
Thrombocytopenia occurs more commonly in patients with eclampsia
(30%) than in patients with both mild and severe forms of preeclampsia
(15%–18%). Of the patients who have severe preeclampsia, 4% to 12% will
manifest criteria for HELLP syndrome (hemolysis,
elevated liver enzymes, and low platelet counts).[9]
Thrombocytopenia was also associated with a higher incidence of preterm
delivery and intrauterine growth retardation. It was concluded that
thrombocytopenia is an independent and important risk factor for the occurrence
of maternal and perinatal complications in PIH.[10] A study on the variation of platelet
function in pregnancy induced hypertension and gestational diabetes mellitus
was done by Zhonghua Fu Chan Ke
Za Zhi and concluded that
Platelet activity is enhanced in PIH and GDM. It may play an important role in
the pathogenesis and development of the two diseases.[11]
Thrombocytopenia is one of the important diagnostic criteria for pre-eclampsia. Thrombocytopenia (count below 100x 103/mm3) is
one of maternal indication for delivery in pre-eclampsia.[12] Hence platelet count becomes one of the
important diagnostic tool in assessing the maternal and fetal well-being. The
estimation of platelet count from blood smears must be systematic each time the
automated count is erroneous because even the most expensive and most effective
machine is not able to replace human judgment.[13-15] Obtaining an
accurate platelet count by using an automated hematology analyzer may be
complicated by the presence of particles of similar size and/or light scatter
properties (red cell fragments, microcytic red cells,
apoptotic white blood cell fragments) and by giant platelets and platelet
clumps.[16,17] .Even the most expensive and accurate hematology
analyzers are not designed to eliminate peripheral blood film evaluation, and
microscopic validation of platelet counts is an important component of the
blood smear review.[18] Although platelet count is a daily routine
laboratory test, the estimation techniques seem to have not been validated.
This is due to the fact that the methods of validation of the diagnostic tests
were finalized during the second half of the 20th century and researchers are
tempted to validate the new methods first, especially the less widespread.[19] Even if the manual platelet numeration,
using a counting chamber, remains the technique of reference, it consumes more
time and requires a phase-contrast microscope, which is not always available in
routine laboratories. In addition, it is worth remembering the important risk
of error estimated up to 10-20% by some authors.[20] Mohamed Brahimi et al performed the estimation of platelet count from
a blood smear on the basis of the red cell: platelet ratio and compared with
the automated platelet count. They concluded that this estimation method is
faster, taking only five minutes on average per patient, while demonstrating
good precision.[18] Malok
et al compared two platelet count estimation methodologies for peripheral blood
smears i.e. traditional estimation method, average number of platelets per oil immersion field
multiplied by 20,000 to yield a platelet by the patient’s hemoglobin value in
g/dl and then multiplied by 1,000 to yield platelet count estimation per micro
liter. The agreement between the two manual methodologies with each other and
each method with automated count was assessed. The study found that the
traditional estimation method provided
more agreement with automated counts
than did the alternate estimation method.[21].Platelet estimation by
peripheral smear method which can be done at any set up, with no availability
of sophisticated equipments with only availability of minimal tools like
microscope, glass slides and Leishman’s stain becomes
useful in early screening of a pregnant women and prevent the complications by
managing or referring the women early for the tertiary care and prevent the
complications to occur in high risk pregnancies. Hence it can be useful test to
prevent the complications of PIH like HELLP syndrome and DIC and thus can be
helpful in reducing the maternal and fetal morbidity and mortality
CONCLUSION:
The
result of this study suggest that platelet estimation by peripheral smear
method is a reliable, rapid, easy and economic, it can be done even in rural
setup for early diagnosis of thrombocytopenia in pregnancy, as it is equivalent
to automated method.
REFERENCES:
1.
Levy JA, Murphy LD. Thrombocytopenia in Pregnancy. J Am
Board Fam Pract 2002;15(4):290 –7.
2.
Sibai BM. Hypertension
in pregnancy. In : Gabbe SG,
Niebyl JR, Simpson JL, (eds).
Obstetrics: Normal and Problem of Pregnancies. 3rd Edn.
New York: Churchill Livingstone. 1996. p. 935–91.
3.
Cunningham FG, Gant NF, Leveno
KJ, Gilstrap LC, Hauth JC, Wenstom KD. Hypertensive disorders in pregnancy. In: Seilis A, Noujaim SR, Davis K, (eds). Williams Obstetrics. 21st Edn. New York: McGraw Hill. 2001. p. 567–618.
4.
Malok M, Titchener EH, Bridgers C, Lee BY,
Bamberg R. Comparison of two platelet count estimation methodologies for
peripheral blood smears. Clin Lab Sci
2007; 20(3):154-60.
5.
Lotspeich-Steininer CA. Introduction
to Hemostasis in Clinical Hematology: Principles,
Procedures, Correlations. Philadelphia: Lippincott Williams
andWilkins. 1992. p. 599-611.
6.
Wyrick-Glatzel J, Hugues VC, Harmening D. Routine
Hematology Methods. Philadelphia: FA Davis. 2002. p. 580.
7.
Maedel LB. Examination
of peripheral blood smears. In: Rodak BF, (edi). Hematology: Clinical Principles and Applications. 2nd
Edi. USA : Saunders. 2002. p. 171–83.
8.
Arias F. Haematological disorders
in pregnancy. In: Arias F, (edi). Practical Guide to
High Risk Pregnancy and Delivery. 3rd edi. Delhi:
Elsevier Store. 2009. p. 475-6.
9.
Sullivan CA, Martin JN, Jr. Management of the obstetric
patient with thrombocytopenia. Clin Obstet Gynecol 1995;38(3):521-34.
10.
Romero R, Mazor M, Lockwood CJ, Emamian M, Belanger KP, Hobbins
JC, et al. Clinical significance, prevalence, and natural history thrombocytopenia
in pregnancy-induced hypertension. Am J Perinatol
1989;6(1):32-8.
11.
Yin SM, Li YQ, Xie SF, Ma LP, Wu
YD, Nie DN, et al. Study on the variation of platelet
function in pregnancy induced hypertension and gestational diabetes mellitus” Zhonghua Fu Chan Ke Za Zhi 2005;40(1):25-8.
12.
Lana K, Wagner MD. Diagnosis and Management of
Preeclampsia. American Family Physician 2004;70(12):2317-22.
13.
Bain BJ. Diagnosis from the blood smear. N Engl J Med 2005;353(4):498-507.
14.
Delobel J. Thrombopenies ('exception des purpuras
thrombopeniques idiopathiques
et des purpuras thrombotiques
thrombocytopeniques). In: Laffont
A, Durieux F, (eds).
Encyclopedie Medico Chirurgicale:
Traite d'Hematologie
(sang). Paris, France: Editions Thechnique. 1997. p.
13-20.
15.
Trzeciak MC, Bordet JC.
Exploration de I'hemostase primaire.
In: Laffont A, Durieux F, (eds). Encyclopedie
Medico-Chirurgicale: Hematologie.
Paris, France: Editions Scientiques et Medicales Elsevier SAS. 2002.
p. 13-9.
16.
Charie LA, Harrison P,
Smith CU, Cobb JR, Briggs C, Machin
S. Accuracy in the low platelet count range: a comparison of automated platelet
counts on Beckman Coulter high-volume hematology analyzers with the ISLH/ICSH
platelet reference method. Lab Hematol 2001;7:236-44.
17.
Rowan RM. Platelet counting and the assessment of platelet
function. In: Koepke JA, editor. Practical Laboratory
Hematology. New York, NY: Churchill Livingston. 1991. p. 157-70.
18. Brahimi M, Osmani
S, Arabi A, Enta-Soltan B, Taghezout Z, Elkahili BS, et al.
The estimation of platelet count from a blood smear on the basis of the red
cell: platelet ratio. Turk J Hematol 2009; 26:21-4.
19. Jenicek M, Cleroux
R. Epidemiologie (Principles, Techniques,
Applications). 3rd edi. Quebec, Canada: Edisem Inc. 1983.
20. Jobin F. L’hémostase.
QuĂ ©bec, Paris: Presses UniversitĂ© Laval, Editions Maloine;
1995. p. 496.
21. Malok M, Titchener
EH, Bridgers C, Lee BY, Bamberg R. Comparison of two
platelet count estimation methodologies for peripheral blood smears. Clin Lab Sci 2007; 20(3):154-60.
Received on 24.06.2015 Accepted on 28.07.2015
© Asian Pharma
Press All Right Reserved
Asian
J. Pharm. Ana. 5(3): July- Sept. 2015; Page 139-141
DOI: 10.5958/2231-5675.2015.00022.8