A Simple Ultraviolet Spectrophotometric Method for the Estimation of Docetaxel in Bulk Drug and Formulation

 

Mahawar Sheetal*

Department of Pharmacy, Shri Jagdish Prasad Jhabarmal Tibrewala University, Vidyanagari, Jhunjhunu, Rajasthan, India

*Corresponding Author E-mail: sheetal.mpharmndds@gmail.com

 

ABSTRACT:

The present investigation aimed at development and validation of an analytical method for the estimation of Docetaxel based on Ultra-violet Spectrophotometry. The reported methods of analysis include the chromatographic methods such as HPLC (1-2), RP-HPLC (3-4), LC-MS/MS (18) and HPLC/MS in plasma (5-7). An ion-pair extractive spectrophotometric method has been reported in the literature for the determination of a docetaxel (8). Chromatographic methods are complex, tedious, time consuming and are not suitable for routine analysis. However, Ultraviolet spectrophotometric methods which are simple and fast are ideal for routine analysis. Docetaxel exhibits a sharp peak at 229nm when scanned in the Ultraviolet region. Thus, it was selected as the analytical wave-length. The developed method was found to be simple, sensitive, accurate, precise and specific as evidenced by non-interference from the excipients present in the marketed injection. Linearity was observed in the range of 5μg/ml to 45μg/ml and Accuracy was found between 99.36 to 101.75%. The limit of detection (LOD) and limit of quantitation (LOQ) were found to be 0.174μ/ml & 0.194μ/ml respectively.

 

KEYWORDS: Docetaxel, Validation, UV spectroscopic method

 


 

INTRODUCTION:

Docetaxel is a semi-synthetic drug, belonging to taxoid family, derived from precursor extracted from the renewable needle biomass of the European yew tree, Taxus baccata. Docetaxel, which is chemically (2R,3S)-N-carboxy-3-phenylisoserine,N-tert-butyl ester, 13-ester with 5β-20-epoxy-1,2α,4,7β,10β,13α-hexahydroxytax-11-en-9-one 4-acetate 2-benzoate, trihydrate, an anti neoplastic agent most commonly used in the treatment of breast, ovarian, prostate, and non-small cell lung cancer (9). The molecular empirical formula for Docetaxel is C43H53NO14.3H2O (10). The structure of Docetaxel is shown in Figure 1. It is highly lipophilic and practically insoluble in water, soluble in ethanol, methanol, chloroform, acetone, acetonitrile (10).

 

It is available in the market as an injection to be administered intravenously. Docetaxel is official in USP. The methods of estimation of Docetaxel in bulk drug and the formulation include the chromatographic methods using UV detector (9). Other reported methods of analysis are reverse phase (3-4) and ion pair (5) HPLC methods. So far, no Ultraviolet spectrophotometric method for estimation of Docetaxel is reported. Ultraviolet spectrophotometric methods are simpler and faster compared to the chromatographic methods. Docetaxel in acetonitrile exhibits a sharp peak at 229 nm when scanned in the UV region between 200-400 nm and hence it was selected as the analytical wave-length. Thus, the aim of the present investigation was to develop a reliable spectrophotometric procedure for estimation of docetaxel in bulk and marketed formulation.

 

Docetaxel was obtained as a gift sample from RPG Life Sciences, Mumbai, India. Acetonitrile AR Grade was obtained from Fischer Scientific, Mumbai, India.  Docetaxel injection IP (20 mg) (marketed formulation) of docetaxel was purchased from a local retail pharmacy.

 


 

Figure 1 Structure of Docetaxel

 

 


Equipment

Digital analytical balance (Shimadzu SCS, Switzerland), Spectrophotometric measurements were made on a Shimadzu 1700 double beam UV-VIS spectrophotometer with a fix slit width of 1nm coupled with computer loaded with Shimadzu UV PC software (UV probe) version 2.31.

 

Methods

Determination λmax

The absorption maximum (λmax) of docetaxel was determined by scanning 10 µg/ml solution against Acetonitrile as reagent blank in spectrum mode between 200 to 400nm.

 

Preparation of Stock Solution

Primary stock solution of docetaxel in acetonitrile (100µg/ml) was prepared by dissolving 5mg docetaxel in 50ml acetonitrile. The primary stock solution was stored at 2-8şC.

 

Preparation of Calibration Curve in Acetonitrile

Appropriate aliquots of the stock solution of docetaxel (100µg/ml) were transferred to 10ml volumetric flasks and were diluted up to the mark with acetonitrile. The absorbance of all the prepared solutions (5, 10, 15, 20, 25, 30, 35, 40 and 45 µg/ml) was then measured at the absorbance maxima, 229nm against the reagent blank (acetonitrile). The readings were recorded in triplicate. Mean value (n=3) along with the standard deviation (SD) are recorded in Table 1. The regressed values of absorption were plotted graphically against the concentrations, as shown in Figure 2. Stability of the solutions of docetaxel in acetonitrile was ascertained by observing the changes in the absorbance of the solution at the analytical wavelength, over a period of 24 hours, at room temperature. The readings were recorded in triplicate.

 

Analytical Method Validation

Linearity

The linearity of an analytical method is its ability to elicit, test results that are directly or a well defined mathematical transformation proportional to the concentration of analytes in samples within a given range (11). Beer’s law states that absorbance is proportional to the concentration of the absorbing species.       A calibration curve is prepared by plotting a dependent variable (absorbance Y) as a function of an independent variable (concentration X). This relation if found with a series of measurements, which in practice, is often a linear one.         

Y= mX +C

Where, m is the slope of the line and C is the intercept on the Y axis (11).

 

Linearity of an analytical method for docetaxel in acetonitrile was established by the regression coefficient.

 

Accuracy

Accuracy of an analytical method is the closeness of the test results obtained by that method to true value (12). Accuracy, sometimes also referred to as recovery is an indicator of the trueness of test measurements. To determine the accuracy of the method three quality control samples (10μg/ml, 20μg/ml and 30μg/ml) for acetonitrile were used. The samples chosen were such to represent the entire range of the standard curve i.e lower, middle and higher concentrations of the range. Accuracy was calculated by analysis of 3 replicate samples for the above described methods. The observed concentrations of the drug were then back calculated using the equation of standard calibration curve and compared with actual concentrations (13-14). The accuracy results for the UV method are listed in table 3. The contents of the vial of the marketed formulation containing 20 mg of Docetaxel were transferred to 100 ml volumetric flask and were dissolved in the acetonitrile. A measured volume of this solution was mixed with standard docetaxel solution and diluted suitably to get the total docetaxel concentrations of 10, 20 and 30 μg/ml respectively. The absorbance of the solutions was measured at 229 nm using acetonitrile as blank. The study was performed in three replicates and the mean was found at each concentration level.

 

Precision

Precision of an analytical method is the degree of agreement among the individual test results when the procedure is applied repeatedly to multiple scanning of homogenous sample (12). Precision may be measure of either degree of reproducibility or of repeatability of the analytical method under normal operating conditions. The precision of an analytical method is usually expressed as the standard deviation or confidence limit. The standard deviation is calculated from the following formula (11).

 

SD = [-1]1/2

Where X = an individual measurement in the set, x=arithmetic mean of the set           

n= total number of replicated measurement taken in set

Precision between different samples can be compared with RSD as follows:

 

% RSD = [SD/Mean]*100

For current assay validation, within day precision and day to day precision were used. Repeatability was determined by using different levels of drug concentrations (5-45µg/ml) for acetonitrile prepared from independent stock solutions and analyzed. Inter-day and intra-day variation was studied to determine precision of the analytical methods. Different levels of drug concentrations in triplicates were prepared three different times in a day and studied for intra-day variation.  Same procedure was followed for three different days to study inter-day variation.  The percent relative standard deviation (%RSD) of the predicted concentrations from the regression equation was taken as precision (13-14). The precision results for the UV method are listed in table 4.

 

Limit of Detection and Limit of Quantification:

The Limit of Detection (LoD) is a quantitative parameter. It is the lowest concentration of the analyte in a sample that can be detected with acceptable precision and accuracy under stated experimental conditions (12). It is expressed as the concentration of the analyte in the sample. The limit is expressed in terms of µg/ml, ng/ml, pg/ml. etc. LoD values are specific for a particular set of experimental conditions. Anything that changes the sensitivity of a method, including instrument, sample preparation etc. will change the detection limits.

 

Limit of Quantification (LoQ) is the lowest concentration of the analyte in a sample that may be measured in a sample matrix such as impurities in bulk drug substances and degradation products in finished pharmaceuticals. The value of LoQ is almost 10 times higher than that of the blank (11). The LOD and LOQ of docetaxel were determined using calibration standards. LOD and LOQ were calculated as 3.3 σ/S and 10 σ/S, respectively, where S is the slope of the calibration curve and σ is the standard deviation of y-intercept of regression equation (n=9) (13-14).

 

RESULTS AND DISCUSSION:

Determination of λmax

Based on the spectrophotometric scanning of docetaxel (10µg/ml), the maxima was obtained at 229nm in acetonitrile, hence chosen as the analytical wavelength.

 

Calibration Curve of Docetaxel

Table 1 shows the mean absorbance values along with the standard deviation of docetaxel in acetonitrile. The high correlation coefficient in the acetonitrile indicated that absorbance and concentration of the drug was linearly related. Beer’s law was found to be obeyed in the range of 5 to 45µg/ml in acetonitrile.

 

Table 1: Absorbance of Docetaxel in Acetonitrile

Concentration

(µg/ml)

Absorbance ± S.D.*

Absorbance ± S.D.*

(after 24 hours)

5

0.113 ± 0.003

0.110 ± 0.013

10

0.221 ± 0.002

0.208 ± 0.018

15

0.324 ± 0.013

0.306 ± 0.020

20

0.433 ± 0.023

0.416 ± 0.020

25

0.530 ± 0.019

0.508 ± 0.015

30

0.645 ± 0.009

0.597 ± 0.012

35

0.749 ± 0.008

0.709 ± 0.009

40

0.851 ± 0.011

0.799 ± 0.010

45

0.960 ± 0.016

0.895 ± 0.020

*Mean ± SD (n =3)

 

Figure 2 Calibration curve of docetaxel

 

Analytical method validation

Linearity of the Assay

 

The linearity of the assay was determined by plotting the standard calibration curves for the concentration range 5-45µg/ml at 229nm in acetonitrile for 3 consecutive days as shown in table 2.

 


Table 2 Linearity of developed analytical method

Solvent

λ max (nm)

Range

Regression Equation

Correlation Coefficient (R2)

Acetonitrile

229

5-45 µg/ml

y= 0.019x + 0.009

0.999

 

Accuracy

The excellent mean % Accuracy values, close to 100%, and their low standard deviation values (SD < 1.0) represent high accuracy of the analytical methods (13-14).

 

The mean % accuracy for lower (10µg/ml), intermediate (20µg/ml), and higher concentrations (30µg/ml), were found to be 99.36 (0.568), 101.75 (0.548) and 99.76 (0.268), respectively in acetonitrile. Thus, the accuracy of the developed method of docetaxel in acetonitrile was found between 99.36 to 101.75%.

 

Table 3 Accuracy of the developed method in Acetonitrile

Concentration (µg/ml)

Calculated concentration (µg/ml) (± SD)*

% Accuracy

(± SD)*

10

9.936 ± 0.046

99.36 ± 0.568

20

20.351 ± 0.089

101.75 ± 0.548

30

29.930 ± 0.066

99.76 ± 0.268

*Mean ± SD (n =3)

 

Above mentioned revealed that any small change in the drug concentration in the solutions could be accurately determined by the proposed analytical methods. The results suggest that the methods were very accurate.                            

 

Precision

Precision was determined by studying the repeatability and intermediate precision. The intra- and inter- day precision of the method are recorded in the Table 4.

 

In precision study, %RSD values were not more than 2.0% in all the cases. RSD values found for the analytical methods were well within the acceptable range indicating that these methods have excellent precision. Precision of the method to estimate docetaxel in presence of the excipients present in the injection was ascertained by performing the assay of the commercially available injection. Briefly, the contents of the vial of the marketed formulation were dissolved in 100 ml acetonitrile. Further, 1.0 ml of the solution was diluted to 100 ml with acetonitrile and absorbance was measured at 229 nm using acetonitrile as blank. The amount of docetaxel present in the formulation was calculated by using the equation generated by the linearity studies as shown in table 5. The above procedure was repeated six times and the standard deviation was calculated for determining intraday precision. The same procedure was repeated six times each on two separate days to determine the interday precision.

 

LOD and LOQ

The LOD and LOQ of docetaxel were determined using calibration standards. LOD and LOQ were calculated as 3.3 σ/S and 10 σ/S, respectively, where S is the slope of the calibration curve and σ is the standard deviation of y-intercept of regression equation (n=9).

 

 


Table 4 Intra day and Inter day precision for docetaxel determination in Acetonitrile

Conc. (µg/ml)

Intra day Absorbance

Inter day Absorbance

Mean ± SD*

% RSD

Mean ± SD*

% RSD

05

0.112±0.0008

0.729

0.112±0.0008

0.729

10

0.204±0.0036

1.767

0.202±0.0040

1.722

15

0.310±0.0035

1.134

0.302±0.0075

1.325

20

0.413±0.0025

0.609

0.409±0.0065

1.837

25

0.508±0.0055

1.085

0.501±0.0072

1.298

30

0.595±0.0044

0.733

0.589±0.0040

1.224

35

0.710±0.0050

0.709

0.705±0.0056

0.574

40

0.795±0.0069

0.871

0.793±0.0055

0.702

45

0.896±0.0056

0.621

0.889±0.0051

0.577

*Mean ± SD (n =3)

 

 

 Table 5. Results of the intra day and inter day precision

Set No.

Assay (% of labeled amount)

Day 1

Day 2

Day 3

1

99.78

99.45

98.23

2

99.86

99.56

99.41

3

99.75

99.25

99.23

4

99.35

98.31

99.64

5

100.32

100.54

98.67

6

101.33

98.54

100.25

 Mean

100.06

99.27

99.23

Standard deviation

0.69

0.80

0.71

 

CONCLUSION:

The UV-visible spectrum obtained by scanning the 10μg/ml of docetaxel recorded between 200 nm to 400 nm. It was observed that docetaxel shows the characteristic peak at 229nm and thus it was selected as the analytical wavelength. Analytical method for docetaxel was developed and validated by UV spectrophotometrically in acetonitrile. The high correlation coefficient in the above solvent indicated that absorbance and concentration of the drug was linearly related. Beer’s law was found to be obeyed in the range of 5 to 45µg/ml for docetaxel in acetonitrile.

 

There is negligible difference in the absorbance values of the fresh and the stored solutions indicating that docetaxel is stable over the period of analysis. The accuracy of the developed method of docetaxel in acetonitrile was found to close to 100%, between 98.00 to 102.00%. In precision study, %RSD values were not more than 2.0% in all the cases. RSD values found for the analytical methods were well within the acceptable range indicating that these methods have excellent precision. The LoD and LoQ for the assay of Docetaxel using Acetonitrile is 0.174 and 0.194 respectively indicating that the method is sensitive. It is also selective as evidenced by the noninterference of the excipients used in the formulation of the injection. The developed spectrophotometric methods for determination of Docetaxel are simple, specific, accurate, precise, rapid and economical which indicates its adequacy for routine pharmaceutical analysis. It is concluded that the developed spectrophotometric method can be successfully utilized for the routine estimation of Docetaxel in bulk as well as formulation.

 

REFERENCES:

1.        Anders Andersen, David J Warren, High sensitivity assays for docetaxel and paclitaxel in plasma using solid-phase extraction and high-performance liquid chromatography with UV detection, BMC Clinical Pharmacology 2006, 6:2.

2.        Annarapu Malleswara Reddy, Nagaraju Banda, Shinde Govind Dagdu, Evaluation of the Pharmaceutical Quality of Docetaxel Injection Using New Stability Indicating Chromatographic Methods for Assay and Impurities Sci Pharm. 2010; 78: 215–231

3.        J.V.L.N. Seshagiri Rao, Sk. Mastanamma, P. Prahlad A validated RP-HPLC method for the estimation of Docetaxel in injectable dosage forms International Journal of Research in Pharmaceutical and Biomedical Sciences Vol. 2 (1) Jan – Mar 2011

4.        Vinay Kumar Venishetty, Nirali Parikh, Application of Validated RP-HPLC Method for Simultaneous Determination of Docetaxel and Ketoconazole in Solid Lipid Nanoparticles India Journal of Chromatographic Science, Vol. 49, February 2011

5.        A.M.Reddy, B.Nagaraju, D.V.Rao, K.C.Sekher, Sci Pharm.,2010,78(2),215-231

6.        S.D. Baker, M. Zhao, P .He, M.A. Carducci, J .Verweij, , Analytical Biochemistry 2004,324,276-284

7.        W.J Loos, J. Verweij, K. Nooter, G. Stoter, A. Sparreboom,. J Chromatogr B Biomed Sci Appl.,1997,693,437-441

8.        Neelam Sridevi , Gopisetti Jahnavi, Chandra Bala sekaran, Der Pharmacia Lettre., 2012,4(1),159-169

9.        T. Naganjaneyulu, T. Archana Devi, Extractive spectrophotometric methods for the determination of docetaxel in pure and pharmaceutical formulations Der Pharma Chemica, 2013, 5(1):131-136 

10.     Noha Nafee, Sebastian Taetz, Chitosan-coated PLGA nanoparticles for DNA/RNA delivery: effect of the formulation parameters on complexation and transfection of antisense oligonucleotides Nanomedicine: Nanotechnology, Biology, and Medicine 3 (2007) 173–183

11.     Rifino C.B. “Pharmaceutical Process Validation”, In: Robert AN, Wachter AH (Eds.) Process Validation and Quality Assurance. Switzerland: Marcel Dekker, Inc.,1-43, (2003) The United States Pharmacopoeia 27, NF 22, United States Pharmacopoeial convention, Rockville, MD, 2004

12.     The United States Pharmacopoeia 27, NF 22, United States Pharmacopoeial convention, Rockville, MD, 2004

13.     J.C. Miller, J.N. Miller, Statistics for Analytical Chemistry, 2nd ed.,Wiley, New York, 1984, pp. 83–117

14.     S.K. Sharma, “Validation of Pharmaceutical Products and Process”. The Eastern Pharmacist, July,21-23(2001)

15.     Bissery, M.C., Guenard, D., Gueritte-Voegelein, F. and Lavelle, F.: Experimental antitumor activity of taxotere (RP 56976, NSC 628503), a taxol analogue. Cancer Res., 51, 4845–4852 (1991)

16.     Lavelle, F., Bissery, M.C., Combeau, C., Riou, J.F., Vrignaud, P. and Andre, S.: Preclinical evaluation of docetaxel (Taxotere). Semin. Oncol., 22, 3–16 (1995) 

17.     Fossella, F.V.: Overview of docetaxel (Taxotere) in the treatment of non-small cell lung cancer. Semin. Oncol., 26, 4–8 (1999)

18.     J. Ciccolini , J. Catalin , M.F. Blachon , A. Durand, Rapid high performance liquid chromatographic determination of docetaxel (Taxotere) in plasma using liquid–liquid extraction, Journal of Chromatography B, 759 (2001) 299–306

 

 

Received on 18.04.2013          Accepted on 26.05.2013        

© Asian Pharma Press All Right Reserved

Asian J. Pharm. Ana. 3(2): April- June 2013; Page 48-52